Sysmex XQ320


1. Clone/Copy the xq320 driver from: https://github.com/HISMalawi/mlab_drivers 
onto the server
2. Navigate to xq320 directory and do: npm install .Since servers do not have internet, 
you can do this on your local machine, make sure node version is the same as that 
on the server via node –version then copy node_modules to this directory on the 
server.
3. Navigate to config/settings.json file of xq320 and change serverIp to your server ip 
address and localhost on lispath to actual server IP. LisUser and lisPassword to 
credentials used to log into iBLIS and serverPort to 5555
4. Open the file xq320.js via vim xq320.js and the last line as follows: where there is 
settings.serverPort, it should be 5555 and where there is settings.serverIP should be 
server ip address just the same as those you have set in config/settings.json
5. Do run this command to allow firewall on the server: sudo ufw allow 5555/tcp
6. In the xq320 folder, run this to start the driver: node xq320.js This should start the 
service and a successful start will show a log that it has started on this ip address and 
port which are set in the config/settings.json.
7. Go to the machine and give the machine network by connecting to it at the back via 
ethernet cable. Make sure the network is capable of DHCP as the machine 
automatically picks the IP address, there is no manual setting of network.
8. On the machine software interface, go to settings then TCP/IP connection and set 
the Host IP as the Server Ip which is the same as that in config/settings.json and the 
port to the port same as serverPort in config/settings.json
9. Save the TCP/IP connections. Upon saving and going back to main screen and clicking 
browser, bottom left of the screen has to show a green bar as shown in the picture 
below. This indicates that there is connection between the service that was run in 
step 4 and the machine. Without this not communication can happen. This also 
shows that network between machine and LAN is Ok via the ethernet.
10. Once this is ok, Register one FBC test in IBLIS, take the accession number without the 
facility code i.e starting from the number, go to the machine, let the lab tech do one 
dummy test but the test should have the sample id as the accession number of the 
just registered test( achieved by entered the accession number on the sample ID 
input field when recording the detail of the sample in the machine before testing it).
11. Once it has finished testing, click on explore on the top navigation on the machine, 
you will be presented with a list as below image. Select the test with the accession 
number you registered(should be highlighted blue), then click browserr. 
 
12. Browser will have the details of the test(results and the sample id i.e accession 
number ) as shown in the picture in step 7. Click validate and once its validated, click 
on Output the HC Outpus. This should send results to the service and you can 
observe the log of the service running that was done in step 4.
13. If all is good, fetch the result in IBLIS for the test you registered and compare the 
results if the machine what is in the browser of step 9.
14. Should some results be missing, check the mapping in config/xnseries.json.
15. If every this is OK, make the service to auto start by doing: pm2 start xq320.js then 
pm2 save.
16. To make the machine automatically sent the results once the machine finishes 
testing, do step 15 and 16.
17. Making sure the test is auto validated once testing finishes by going to the machine 
interface Menu, then settings and choose the settings for auto validation (can be 
found in operational) as below image. Then click modify settings and set auto 
validation to be for all samples and save.
18. Making sure the test results are automatically sent to iblis once testing finishes by 
going to the machine interface Menu, then settings and choose the settings for auto 
output (can be found in operational) as below image. Then click modify settings and 
make sure HC is select to as well to be an auto output destination as well and save.
That’s it. You can try testing an actual sample and verify if its sent to iblis automatically 
without you clicking on output as was done in step 10.